This tab is used for setting parameters such as Energy, Attenuation, Beam size, Detector Distance, Detector 2Ө
(to be implemented), Beamstop position, and Goniometer Omega.
Below is a screenshot of the Beamline Setup tab.

The functionality of each button is described
below.
Prepare for Mounting:

The Prepare for Mounting button
is used to prepare the hutch for mounting crystals. This button
will move the detector and beamstop out of the way to make space
to mount your crystal. After mounting the crystal, you will
probably need to align within the Sample Camera pane described
below.
Sample Camera and Sample Alignment
The Sample Camera pane shows a live view of the
crystal and has a few simple buttons to help you align the your
crystal.

Zoom Level:

If the crystal is not visible, you may need to
decrease the zoom level using the zoom level buttons.
Move Sample:

To get the crystal in the field of view or for
manual centering, you may used the move sample buttons.
Rotate Sample by:

The sample needs to be rotated as part of the
alignment process to ensure the axis of rotation of the crystal
intersects the beam. The crystal is aligned by centering the
crystal as described below.
Centering:

Manual point and click centering is activated by
selecting the 'Click' button, within the Sample Camera
tab in the MX Data Collection GUI. The crystal is aligned
as follows:
- Select a low Zoom Level to find the
sample.
- Activate 'Click' Centering and Click
on the sample to center it.
- Rotate Sample By
+90 (or -90) degrees.
- Click on the sample once again to
center.
- Select a high Zoom Level.
- Repeat steps 3 to 4 to align the crystal
accurately.
After aligning the crystal you should check
the beamline parameters and adjust them if necessary.
Beamline Parameters

The beamline parameters shown above (Energy, Attenuation, Beam
Width, Beam Height, Goniometer Omega, Beamstop position, Detector Distance,
and Detector 2Ө
(to be implemented), can be adjusted by entering the desired
value in the editable box on the left (the box on the right
shows the current value). After entering the value, you must
either hit ENTER or click the right-arrow to apply the
value. While the parameter changes, you will see its value
updated in the current value box, and the right-arrow will turn
into a stop button that you may use to cancel a parameter
change. The curved left-arrow may be used to revert to a
previous value.
Energy (keV):

The Energy box is used to select the energy of
photons that reach the sample in the endstation. This is
achieved by moving the monochromator to the appropriate Bragg
angle. This box is useful for setting the energy that you wish
to take a snapshot of the crystal at.
Attenuation (%):

The Attenuation box is used to control the
percentage of photon that are transmitted to the sample in the
endstation, and hence the intensity of diffraction spots on the
detector. If the diffraction spots are too intense, you may
either increase the attenuation or decrease the exposure time.
The attenuator has four thin foils of aluminium that are set
according to the desired attenuation. The current value of the
attenuation is calculated based (with beam energy taken into
account) on the actual foils used.
Beam width (mm):

The Beam width box allows the user to control
the width of the JJ slit collimators that trim the beam in the
horizontal direction.
Beam height (mm):

The Beam width box allows the user to control
the width of the JJ slit collimators that trim the beam in the
horizontal direction.
Omega (deg):

The Omega (deg) box allows the user to control
the rotation angle of the sample goniometer.
Beam-stop (mm):

The Beam-stop (mm) box allows the user to
control the distance of the beamstop from the sample (it can
approach to within about 5mm and extend out to about 45mm). The
further the beamstop is from the sample, the larger its shadow
will be on the detector.
Detector Distance (mm):

The Detector Distance (mm) box allows the user
to control the distance of the detector from the sample. The
detector distance may also be set from within the Data
Collection tab, but the value entered hear will become the
default value.
Detector TwoTheta:

This feature is not available directly from
the GUI at the moment, but the detector can be put into the
appropriate position